FDA Recalls Subcutaneous-ICD Leads Due to Fracture Risk

The US Meals and Drug Administration (FDA) has issued a Class I recall of the electrode element of the EMBLEM S-ICD (Boston Scientific), the long-lasting subcutaneous implantable cardioverter-defibrillator system, as a consequence of an elevated danger for fracture at a selected location close to the proximal sensing ring.

Such a lead failure may forestall the machine from delivering shocks as wanted, the company stated in its recall announcement, which applies to 19,919 mannequin quantity 3501 gadgets manufactured from March 2016 to November 2020 and distributed till November.

“There have been 27 complaints about this machine subject and 26 experiences of significant accidents. One loss of life has been reported,” the announcement stated.

Boston Scientific had issued a security advisory on December 2 “to all affected clients, with suggestions for immediate identification of gadgets in danger for electrode physique fracture.” The notice included directions for monitoring machine perform each remotely and radiographically, and for substitute of leads displaying proof of a fracture.